60601-1 © IEC:2005
60601-1 © IEC:2005
– 337 –
– 169 –
12.4.5.4 Other ME EQUIPMENT producing diagnostic or therapeutic radiation
When applicable, the MANUFACTURER shall address in the RISK MANAGEMENT PROCESS the
RISKS associated with ME EQUIPMENT producing diagnostic or therapeutic radiation other than
for diagnostic X-rays and radiotherapy (see 12.4.5.2 and 12.4.5.3).
Compliance is checked by inspection of the RISK MANAGEMENT FILE.
12.4.6 Diagnostic or therapeutic acoustic pressure
When applicable, the MANUFACTURER shall address in the RISK MANAGEMENT PROCESS the
RISKS associated with diagnostic or therapeutic acoustic pressure.
Compliance is checked by inspection of the RISK MANAGEMENT FILE.
13 * HAZARDOUS SITUATIONS and fault conditions
13.1 Specific HAZARDOUS SITUATIONS
* General
13.1.1
When applying the SINGLE FAULT CONDITIONS as described in 4.7 and listed in 13.2, one at a
time, none of the HAZARDOUS SITUATIONS in 13.1.2 to 13.1.4 (inclusive) shall occur in the
ME EQUIPMENT.
The failure of any one component at a time, which could result in a HAZARDOUS SITUATION, is
described in 4.7.
* Emissions, deformation of ENCLOSURE or exceeding maximum temperature
13.1.2
The following HAZARDOUS SITUATIONS shall not occur:
– emission of flames, molten metal, poisonous or ignitable substance in hazardous
quantities;
– deformation of ENCLOSURES to such an extent that compliance with 15.3.1 is impaired;
–
temperatures of APPLIED PARTS exceeding the allowed values identified in Table 24 when
measured as described in 11.1.3;
temperatures of ME EQUIPMENT parts that are not APPLIED PARTS but are likely to be
touched, exceeding the allowable values in Table 23 when measured and adjusted as
described in 11.1.3;
–
– exceeding the allowable values for “other components and materials” identified in Table 22
times 1,5 minus 12,5 °C. Limits for windings are found in Table 26, Table 27 and Table 31.
In all other cases, the allowable values of Table 22 apply.
Temperatures shall be measured using the method described in 11.1.3.
The SINGLE FAULT CONDITIONS in 4.7, 8.1 b), 8.7.2 and 13.2.2, with regard to the emission of
flames, molten metal or ignitable substances, shall not be applied to parts and components
where:
– The construction or the supply circuit limits the power dissipation in SINGLE FAULT
CONDITION to less than 15 W or the energy dissipation to less than 900 J. '